Order information Index Date of Issue SYSTEM PERFORMANCE The procedures described in this manual must be observed in order to ensure proper system performance, and to avoid hazards. Table of Contents ABL80 FLEX analyzer Operators manual 1. Impact of the pO2 level on Glucose linearity If the pO2 level in a sample is: Technical documentation Then cGlu linearity specifications only apply to cGlu values between: Linearity not specified.Ī. Please note that glucose performance is not specified when the pO2 is less than 20 mmHg (2.7 kPa). Specification CAUTION: Low pO2 levels can influence the linearity of glucose measurements, and can therefore result in falsely low glucose results. This information is already provided in the ABL80 FLEX Reference Manual, chapter 6. Instructions to user Please place this document in the binder of your manual.īrief Overview of the changes Changes Description To increase operator awareness, the specifications listed here regarding glucose linearity in the ABL80 FLEX analyzer will be added to the operators manual. The following table outlines the glucose linearity as a function of the pO2. Low pO2 levels can influence the linearity of the glucose sensor.